MRF 16/22
44
MF 55-GB 1.21
13.5 Fuse sizes and fuse replacement
Access to fuses for the heating elements and supply to the microwave circuit is by removal
of the left side cover
[See figure 13.31].
Internal Fuses
Power
Supply
Phase
Fuses
Aux Fuse
Line Fuses
Thyristor
Fuses
Microwave
S.C.
HT Fuses
F1, F2, F3
F4, F5
F6, F7
F8, F9,
F10, F11
F14
3 x 40A
2 x 5A
2 x 25A
4 x 30A
1 x 20A
14 x 51mm 5 x 20mm 10 x 38mm 10 x 38mm 10 x 38mm
208 -
240V
3 Phase
Delta
Type GG
Type F
Type GG
Protistor
Type GG
F15, F16
2 x HVF10
F1, F2, F3
F4, F5
F8
F14
3 x 25A
2 x 5A
1 X 30A
1 x 20A
10 x 38mm 5 x 20mm
10 x 38mm 10 x 38mm
380 -
415V
3 Phase
plus
Neutral
(STAR)
Type GG
Type F
N/A
Protistor
Type GG
F15, F16
2 x HVF10
To access fuses F1 to F13 isolate the furnace from the electrical supply and remove the left
side cover
. [See figure 13.1].
To access fuses F14 to F16 isolate the furnace from the power electrical power supply and
remove the right side access panel followed by the removal of the High Voltage Circuits
cover. Note that the High Voltage Circuit could have energy stored in the capacitors, so
ensure that they are discharged before replacing any fuses.
Fig 14.2 Internal fuses