13
2. Carefully position forks to lift unit from rear. Ensure unit is fully positioned on fork truck prior to raising the
forks.
3. Raise the unit approximately six inches to allow access to the feet and casters and pull out the pallet.
Person 1:
Stand at the front of the unit and provide additional support to the unit.
Person 2:
If feet are desired, lower all 4 factory installed feet (using the 3/8" wrench to loosen if necessary).
If rolling casters are desired, remove the feet and install the 4 casters and tighten using wrench.
Refer to
Without Forklift
section (step 5, 6 and 7) for tightening the casters.
Chromatography Refrigerators
The chromatography refrigerator is equipped with a safety circuit. When the safety circuit is tripped, the
safety sensor senses a flammable gas and the power to the internal outlet will be discontinued. The user will
need to manually reset the safety circuit assembly using the reset switch at the right side of header panel
before the power is restored. To do this, toggle the switch located on the top right side of the header panel
beside the remote alarm contacts. There will be a power delay to the outlet for approximately 4 minutes
where the red LED will be illuminated.
When there is power at the outlet, the red LED will turn off and the green LED illuminates.
If the circuit trips from RED to GREEN repeatedly or never turns GREEN, contact Customer Service.
Flammable materials should not be stored in the refrigerator.
Lights
The GFCI outlet assembly is located inside the unit on the rear wall at the center.
The outlet assembly has two LEDs:
• RED (on the top) - When the red LED is ON, it indicates there is no power to the AC outlet inside the unit.
• GREEN (on the bottom) - When the green LED is ON, it indicates there is power to the AC outlet inside the
unit.