6
CONFIGURATION, CONT.
GRAVITY COMPENSATION
1. With the display showing
GrAu
:
To Enable the Gravity Compensation mode, press the SET key. The display will
change to
On
(use the factory set value of 9.797).
To Disable the Gravity Compensation mode, press the UNIT key. The display will
change to
OFF
(use the value obtained from the factory Tech Support).
2. Press the SET key again. The display will change to show
GrAu
.
3. Press the UNIT key. The display will change to show
End
.
4. Press the SET key to save the settings and reset the scale.
5. Setup is complete.
GRAVITY COMPENSATION
The Gravity Compensation for your scale has been pre-configured at the factory and should not
require changing in most locations. In the event that the factory settings do not meet the
requirements of your location, contact the factory Tech Support for the value for your location.
With
GrAu
enabled in the function setting of the Configuration mode, the scale will use the
factory set value of 9.797 for the Gravity Compensation.
If the
GrAu
is disabled, the scale will use the value obtained from the factory Tech Support
and entered by the operator for the Gravity Compensation.
To change the Gravity Compensation value:
1. Press and hold the UNIT key while pressing the ON/OFF key to turn on the scale and
enter Gravity Compensation mode, the display will show 9.797 (or the current value).
2. Press the UNIT key to increase the value or TARE key to decrease the value to achieve
the desired value.
3. Press the SET key to confirm the entered gravity and return the scale to the normal
operating mode.
CALIBRATION PROCEDURE
1. While the scale is off, press and hold the ON/OFF key.
2. Once all display segments are on, release the ON/OFF key and immediately and press
the UNIT key three (3) times.
3. The scale will show “11111”
”55555”, and finally show the counts of zero.
4. Press the SET key. The scale will show “0g” and then go back to the counts.
5. Place a 5 kg test weight on the scale and then press the SET key.
6. The scale will show “5000g”, and change to show “CAL”, and then automatically turn off.
7. Remove the test weight from the scale.
8. The scale is ready for normal operation.