User Manual: xr100+_eng
31
Cardioline S.p.A.
Via Linz,151
38121 Trento (TN)
info@cardioline.it
www.cardioline.com
Sales office:
Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470
fax +39 02 94750471
xr100+
8.2.4 Inserting the Microphone
Slip the microphone into the pocket, the metal side facing
the arm.
Guide the microphone cable out of the pocket and to the
right of the Velcro tab. Then close the tab.
Fold the end of the cuff over and introduce it into the
metal clasp.
Figure 8 – 2: Inserting the Microphone
8.3 Maintenance
8.3.1 Checks before each use
Before each use, visually inspect the device for signs of
damage. If you detect damage or impaired functions which
may result in a hazard to the patient or the operator, the
device must be repaired before it can be used again.
8.3.2 Technical Safety Inspections,
Inspections of the Measuring
System
The technical safety inspections and the inspections of
the measuring system must be completed every two years
according to the rules of the art by a Service Engineer
authorized by ergoline.
Similarly, the automatic sphygmomanometer in the control
terminal must be checked and, if necessary, calibrated by
an authorized specialist every two years to fulfill legal
requirements.
The date of the next inspection is indicated on the inspection
sticker attached next to the type plate on the ergometer.
8.4 Disposal
The product described in this operator manual must not
be disposed of as unsorted municipal waste and must be
collected separately.
Please contact your authorized manufacturer ergoline GmbH
for information concerning the disposal of your equipment.
There is no waste approval. Proper disposal is documented
by ergoline GmbH.
Consult Operator’s Manual!