9 Technical Specifi cations
9.1 Ergometer
Model
modular ergometer system xr100
models xr100+ and
Operating mode
continuous operation
Power
100 – 240 V / 50 – 60 Hz
max. 100 VA (max. 140 VA for
ergometers with motor-assisted
adjustment of the handlebar)
specifi cations of the US power cord:
SPT 2x18AWG 125 V / 10 A
“hospital” or “hospital grade”
specifi cations of the internal backup battery:
IEC: CR 2032 / 3 V 230 mAh
Braking principle
computer-controlled eddy current brake with torque mea-
surement; speed independent to DIN VDE 0750-0238
Load range
6 … 999 Watt, speed (RPM)-independent
(see diagrams on page 35)
Speed range
30 … 130 n/min
Load accuracy
to DIN VDE 0750-0238
Load increments
user programmable
Internal protocols
control terminal P:
• 5
fi xed and 10 user-programmable protocols
• manual load control
Permitted patient weight
version 1:
manual adjustment (standard), up to 160 kg
version 2:
gas-spring assisted adjustment of the saddle
height (option), up to 200 kg
version 3:
electrical adjustment of the saddle height with
digital indication of the current saddle height
(option), up to 200 kg
Handlebar adjustment
for patient heights between 120 cm and 210 cm
continuous handlebar adjustment over 350°
Handlebar height adjustment
rigid steering column
Crank length
170 mm (adjustable length cranks available as optional
accessories)