CuroCell
®
M4
10
4.15
Table of notifications
Information about each notification is shown in the notification table:
•
Notifications from 1-7 are both audible and visual. The error code will be displayed until the
error has been rectified. If the mute button is pressed the audible warning will crease for a
period of 5 minutes and will return until the error has been rectified.
•
The notifications 8-10 have no audible alarm. The error code is shown until the system is
restarted.
If any of these problems remain, contact technical support.
Notifiering
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the
control unit is in direct sunlight, relocate it.
2
Default settings are not completed, contact technical support.
3
Incorrect input voltage. Make sure that the correct power
supply is used.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter.
5
High pressure. The pressure cannot be reduced to the desired
value within the time limit.
6
The mattress control parameters have not been read. Connect
the CPR.
7
The mattress control parameters have been changed during the
use. Restart the system.
8
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes.
Information to service technicians: this notification shows a
leakage in the blue cell section. More information in the
Servicemanual for CuroCell® M4.
9
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes.
Information to service technicians: this notification shows a
leakage in the green cell section. More information in the
Servicemanual for CuroCell® M4.
10
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes.
Information to service technicians: this notification shows a
leakage in the red cell section. More information in the
Servicemanual for CuroCell® M4.