CuroCell
®
A4
15
Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the
control unit is in direct sunlight, relocate it.
2
Default settings are not completed.
3
Incorrect input voltage. Make sure that the correct power supply
is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter.
5
Automatic setting failure. The correct pressure has not been
reached within the time limit.
6
Leakage under the automatic setting period. The mattress is
leaking too much for the weighing to be accomplished. Control the
mattress and the air connections.
7
High pressure. The pressure cannot be reduced to the desired
value within the time limit.
8
The automatic setting has been restarted too many times during
the automatic setting period.
9
The mattress control parameters have not been read. Connect the
CPR.
10
The mattress control parameters have been changed during the
use. Restart the system.
11
Leakage in one of the sections. Secure the CPR, mattress, and
connection tubes. Information to service technicians: this
notification shows a leakage in the blue cell section. More
information in the Service manual for CuroCell® A4.
12
Leakage in one of the sections. Secure the CPR, mattress, and
connection tubes. Information to service technicians: this
notification shows a leakage in the green cell section. More
information in the Service manual for CuroCell® A4.
13
Leakage in one of the sections. Secure the CPR, mattress, and
connection tubes. Information to service technicians: this
notification shows a leakage in the red cell section. More
information in the Service manual for CuroCell® A4.