Operation CuroCell® IQ
Following operations apply only to CuroCell® IQ regardless of which mattress that is used.
Operation CuroCell® IQ (how to get started)
1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1.
2. The mattress starts to inflate. This takes about 20-40 minutes depending on the size of the mattress.
While the mattress is inflating, the “silent information signal” and “information signal” diodes light up in
orange. When these diodes goes out, the user can be placed on the mattress.
3. The control unit sets the inner pressure of the mattress according to the weight, length, and position
of the user. This takes about 20-30 minutes. During this time, the diode above the “check symbol”
flashes. When the diode stops flashing, the mattress is set and the inner pressure has adjusted according
to the user.
4. Perform a hand check to ensure that the settings are correct.
Figure 1
. On/Off switch.
Note!
•
Each time the system starts up, it will operate as
follows:
When using a mattress without an air safety
mattress (CuroCell® OP10, CuroCell® CX10 or
CuroCell® CX16), the control unit will completely
inflate the main cells and then perform an automatic
setting.
•
When using a mattress with an air safety mattress
(CuroCell® CX15 or CuroCell® CX20), the control unit
always begins by completely inflating the safety
mattress, followed by the other cells, and then
perform an automatic setting.
•
During the automatic setting, try to avoid larger
movements on the mattress. Otherwise, you will get
a notification on that the desired value could not be
reached within the time limit and the weighting must
start over.
•
When using a mattress with a safety mattress in
foam (CuroCell® CX16), the user can lie on the
mattress even when not inflated. When using a
mattress with an air safety mattress (CuroCell® CX15
or CuroCell® CX20), this must be inflated before the
user can lie on the mattress. In both cases, the safety
mattress will carry the load.