Table of notifications
Information about each notification is shown in the notification table:
•
Notifications from 1-10 are both audible and visual. The error code will be displayed until the
error has been rectified. If the mute button is pressed the audible warning will crease for a
period of 5 minutes and will return until the error has been rectified.
•
The notifications 11-13 have no audible alarm. The error code is shown until the system is
restarted.
10
The mattress control parameters have been changed during
the use. Restart the system. If the problem remains, contact
the technical support.
11
Leakage in one of the sections. Secure the CPR, mattress, and
connection tubes.
Information to service technicians: this notification shows a
leakage in the blue cell section. More information in the
service manual for CuroCell® IQ.
Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If
the control unit is in direct sunlight, relocate it. Otherwise,
contact the technical support.
2
Default settings are not completed. Contact the technical
support.
3
Incorrect input voltage. Make sure that the correct power
supply is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air
filter. If the problem remains, contact the technical support.
5
Automatic setting failure. The correct pressure has not been
reached within the time limit. If the problem remains, contact
the technical support.
6
Leakage under the automatic setting period. The mattress is
leaking too much for the weighing to be accomplished.
Control the mattress and the air connections. If the problem
remains, contact the technical support.
7
High pressure. The pressure cannot be reduced to the desired
value within the time limit. Contact the technical support.
8
The automatic setting has been restarted too many times
during the automatic setting period. Contact the technical
support.
9
The mattress control parameters have not been read.
Connect the CPR or contact the technical support.