Operation CuroCell® M4
Following operations apply only to CuroCell® M4 regardless of which mattress that is used.
Operation CuroCell® M4 (how to get started)
1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1.
2. Use the weight settings to set a suitable air pressure.
3. The mattress starts to inflate. This takes about 20-40 minutes depending on the size of the
mattress. While the mattress is inflating, the “mute the information signal” and “information signal”
diode light up in orange. When these diodes goes out, the user can be placed on the mattress.
4. The pulsating mode setting is pre-set. If a different setting is desired, it can be selected when the
user is placed on the mattress.
5. When the user is placed on the mattress and the desired program is selected, the inner pressure of
the mattress is adjusted according to the selected weight setting. This takes about 10 minutes. During
this time, the diode above the selected program flashes. When the diode stops flashing, the mattress
is set and the inner pressure has reached the selected value.
6. Perform a hand check to ensure that the settings are correct.
7. If the mattress is too soft or too hard, adjust the weight setting.
8. Perform another hand check.
Figure 1.
On/Off switch.
Note!
•
Each time the system starts up, it will operate
as follows:
When using a mattress without an air safety
mattress (CuroCell® OP10, CuroCell® CX10 or
CuroCell® CX16), the control unit will adjust to
the set weight.
•
When using a mattress with an air safety
mattress (CuroCell® CX15 or CuroCell® CX20),
the control unit always begins by completely
inflating the safety mattress, followed by the
other cells, and then perform an automatic
setting.
•
When using a mattress with a safety mattress
in foam (CuroCell
®
CX16), the user can lie on
the mattress even when not inflated.
•
When using a mattress with an air safety
mattress (CuroCell
®
CX15 or CuroCell
®
CX20),
this must be inflated before the user can lie on
the mattress. In both cases, the safety
mattress will carry the load when the orange
diode goes out.