To replace the air filter:
1. Loosen the small protective plate on the rear of the control unit
using a size T10 Torx screwdriver.
2. Remove the filter from the holder.
3. Place the new filter in the holder with the pink side facing
outwards. Put the protective plate back in place and secure using
the screws.
If the control unit is used in a dirty environment the filter should
be checked regularly.
9
Storage
It is advisable to store the mattress and control unit in the product bag (accessory), original package or
equivalent for protective storage. Handle the packaged product with caution. Do not place any heavy
objects on top of it. For additional information about storage temperature, see section 12.
10
Maintenance
10.1
General
We recommend that the control unit will be regularly serviced and inspected to maintain functionality
and performance.
Service and maintenance must always be performed by Care of Sweden or one of its authorized
technicians. Only use spare parts approved by Care of Sweden. For more information, see the service
manual for CuroCell® M4, A4 or IQ.
10.2
Between users
Between users, the Pack&Go function should be used to reset the system. When starting up the
product again it will be set on the pulsating mode.
Between users, also check that:
1.
The power cable and power supply are undamaged.
2.
The connecting tubes (marked CPR) on the side of the control unit are positioned correctly
and not leaking.
3.
The hygiene cover is intact and the cover and cells are correctly assembled.
4.
No tubes or connectors are damaged or jammed.
Contact Care of Sweden or your local distributor if any spare parts are required.
10.3
Replacing the air filter
Before any maintenance is done, make sure that the control unit is turned off. Services shall not be
done while using the product.