8. Troubleshooting
Pump does not start
Check that the power cable is connected to a socket
.
Check
the fuses.
Notification triggered
Check that the quick connectors for the tubes on the side
= mattress losing air
of the pump are properly connected.
Check that the CPR valve is closed.
Open the mattress and check for any leaks from air cells or
tubes. Check that all tube connectors are fine and
properly
connected
.
No air in head cells
When inflating the mattress the three static head cells are
inflated last.
User bottoms out
Check the mattress in accordance with “Notification
triggered” above.
See sections 4.3 and 4.4 for more.
Mattress not stationary
Check that the mattress is secured with the elasticated
straps to the underlying mattress or, if the mattress has an
integrated foam cushion, secured with the straps to the
bed base.
Some cells less inflated
This is normal as the air supply alternates between cells
every cycle (one cycle = ten minutes).
Noisy pump, vibrations can be
Check that the pump is correctly attached to the bed.
felt
Resonance may occur in parts of the bed. Remove the
pump and listen to see whether there’s any difference. Can
be resolved by placing the pump on a level surface or
putting, for instance, a towel between the pump and the
bed.
If your questions are not answered by the above information, please contact Care of Sweden
or
your local distributor.