Configured Options (Continued)
The following options are only configurable via the Guardrails® Editor Software (PC based), see Guardrails® Editor Directions For Use
(1000PB01398) for details on how to configure Profile Configurations.
Quiet Mode
Mode to silence key press tones and power down sequence.
Audio Volume
The audio alarm volume of the pump (high, medium or low).
Auto Night Mode
Main
Display (Backlight) dims between hours 21:00 and 06:00.
AC Fail
The AC Power Failure Alarm can be set to sound or be silent if the AC power is
disconnected.
Auto Save
Feature to retain previous settings when pump is switched on.
Event Log
The event log can be set to be displayed or not on main display. Events still recorded in
event
log
if
disabled.
Battery Icon
Indicator displaying the remaining estimated battery capacity.
Callback Time
Adjusts the length of time before the pump sounds the call back alarm.
Drug Override Mode
Always
- Any changes made to the dose rate that are outside of the Guardrails® Soft Alerts
will require confirmation before starting infusion.
Smart
- Confirmation of setting will be required on first dose rate set outside of the Guardrails®
Soft Alert. Any subsequent changes will not require confirmation until after the dose rate has been
confirmed inside the Guardrails® Soft Alert limits. Additionally any changes in dose rate from above
a Soft Alert Max to below a Soft Alert Min or from below a Soft Alert Min to above a Soft Alert Max
will also need to be confirmed.
Maximum Pressure
The maximum occlusion pressure alarm level that can be selected during an infusion.
Auto Pressure
Feature to set the occlusion pressure alarm level to an amount (mmHg) above the current in-line
pressure, using a single key press.
Auto Set Pressure
Automatic feature to set the occlusion pressure alarm level to an amount (mmHg) above the current
in-line pressure, 15 minutes after starting the infusion.
Auto Offset
The automatic offset value in mmHg used by auto pressure and auto set pressure.
Back Off
An automatic feature which is activated following an occlusion. The pump action reverses and
pumps backwards to release the pressure which has built up in the infusion system, this minimises
the post occlusion bolus.
Pressure Display
Sets whether the Pressure Information is available on the main display.
Default Weight
The default patient weight in kg.
Maximum Weight
The maximum patient weight in kg. This is a Guardrails® Soft Alert and can be overridden.
Minimum Weight
The minimum patient weight in kg. This is a Guardrails® Soft Alert and can be overridden.
Rate Lock
Anti-tamper feature which prevents rate changes, bolus operations and powering pump down.
Rate Titration
Feature to adjust the infusion rate while the pump is infusing, without putting the pump on hold.
Cap Rate
The maximum value for infusion rate.
Cap Bolus Rate
The maximum value for bolus rate.
Purge Rate
The rate used during purge operation.
Purge Volume Limit
The maximum permissible purge volume.
Purge Syringe
Feature which prompts the user to purge the extension set prior to the start of the infusion.
Manual Bolus
Bolus delivered by manually moving the plunger mechanism during an infusion or while on hold.
Volume infused displayed will be increased accordingly.
KVO
Sets the Keep Vein Open (KVO) rate at which the pump will operate when End of Infusion (EOI) is
reached. Also allows KVO at EOI to be disabled.
Near End of Infusion Point
Sets the Near End Of Infusion warning time, as time left to End Of Infusion.
End of Infusion %
Sets the End Of Infusion point, as a percentage of syringe volume.
VTBI Clear Rate
Infusion rate will be set to zero when VTBI has been completed.
Guardrails® Editor Software Profile Configuration