file: n:\article\cos14310m5-v1_07car-en\20140117_cos14310m5-v1.07car-en_handbuch_carefusion_laufband-ergometer_print.doc
© 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 17.01.2014 printed 07.05.2014 page: 8 of 16
11.2
Treadmill ergometer, 170/65 (LE250, LE300, LE500)
LE300 CE
LE250 CE
LE300 CE
LE500 CE
UserTerminal
(display & keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED.
Power supply AC
(standard, read nameplate),
dedicated line required.
U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~
including N and PE
f: 50/60 Hz
Fuse (standard, read nameplate)
C 16 A at 230 V ~
Power input (long time)
1600 VA
Power input (momentary)
3500 VA
(peak performances for milliseconds; max. 1s/min during heel strike)
Drive motor capacity
3300 VA
Elevation motor capacity
n.a.
500 VA
Safety standards
n
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), VDE 0751,
IEC EN 60601-1-2, IEC EN 60601-1-4,
EN 957-1, EN 957-6, Machinery Directive 2006/42/EC
Leakage current
0.2 mA
Isolation transformer
2000 VA
Degree of protection:
class I
M
/ type B
R
/ IP 20
Overvoltage category
Category II: Life-Earth < 300 V
transient over-voltage limit 2500 V
Mode of operation
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
sports and medical: S,I,A (EN957)
Classification according to MDD
active therapeutic device and active diagnostic device, risk class II b
Speed
Definition
0.0 ... 22.0 km/h
0.1 km/h
0.0 ... 40.0 km/h
0.1 km/h
Speed accuracy*
+/- 5 %
Option special speed
0.0 ... 10.0 km/h, 0.0 ... 25.0 km/h,
0.0 ... 30.0 km/h, 0.0 ... 40.0 km/h
0.0 ... 10.0 km/h
Angle of inclination**
Definition
n.a.
0.0 ... 25.0 %
0.1 %