file: n:\article\cos14310m5-v1_07car-en\20140117_cos14310m5-v1.07car-en_handbuch_carefusion_laufband-ergometer_print.doc
© 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 17.01.2014 printed 07.05.2014 page: 9 of 16
LE300 CE
LE250 CE
LE300 CE
LE500 CE
Dimension of running surface
length x width
1700 x 650 mm
Dimension frame
length x width x height
2300 x 1050 x 1410 mm
(UserTe 150 mm width)
Track access height from floor
230 mm
Weight of device
330 kg
335 kg
Weight load limits to floors
(traffic load to DIN 1055 part 3)
6.2 kN/m²
In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight
200 kg (deviation on request specially made)
Environmental conditions:
Transport and storage
temperature: –30...+50 °C,
humidity: 0...95 % - without condensation
barometric pressure: 700...1060 hPa
Environmental conditions:
Operation
temperature: +10...+40 °C (deviation on request),
humidity: 30...70% - without condensation (up to 95% on request),
barometric pressure: 700...1060 hPa
Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Pollution degree
Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.
Central lubrication system
manual
Number of stored programmes
6 training profiles / 10 test profiles / 8 user definable profiles
RS232 Digital interface /
serial port
(USB-converter optional)
COM1: standard
COM2: optional
COM3: optional (standard for service)
COM4: optional
Heart rate measurement
POLAR, 1 channel receiver, ECG precisely
PC software
(not for medical applications)
h/p/cosmos para control
®
included
options: h/p/cosmos para graphics
®
and h/p/cosmos para analysis
®
Reverse belt rotation
optional
Safety arch with chest belt
system***
optional
Handrails options:
short,
long, crossbar (standard)
Arm support
optional
Additional keyboard
optional
* Overload and weak power supply (under voltage) could lead to higher discrepancies concerning the speed display or dripping of the fuse.
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse,
but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum
elevation of 25 %.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety arch with chest belt system) is necessary and
highly recommended. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.