41
ENG
Control
WTS LARGE +0300131IT rel. 1.0 - 05.09.2022
10.4 Setting options at operator
level
The following factory-set parameters can be changed at
the operator level (password 0077).
Setting the required conductivity:
3B LIMIT OUT
xx,x µS
By pressing the arrow keys
éê
you can set the max.
set point
»
Confirm the selection with
Setting the required flushing time / flush-ing cycle:
11G TIME
00 min 00 sec
By pressing the arrow keys
éê
you enter the duration
of the flushing time for the peri-odic module flush..
»
Confirm the selection with
. The program
jumps to submenu "11H".
11H WASH CYCLE
00 min 15 sec 24 h
By pressing the arrow keys
éê
you specify the time
period between the periodic mod-ule flushes.
e.g. change 24h to 48h (usage pause)
»
Confirm the selection with
.
The program jumps back to the main menu.
Notice: In the factory settings, parameter 11
CLEANING is preset to ON. If this setting has been
changed to OFF in the service level, no entries can be
made here in the operator level.
Setting the required language:
In this submenu you can select the language shown on
the display.
16 LANGUAGE
17 SET PASSWORTD
LANGUAGE
ENGLISH
»
Press the arrow keys
éê
to select your language
»
Confirm the selection with
.
The program jumps back to the main menu.