In expert mode, move the bladder image to the center. Dark green central line indicates the image is
centered, while the orange central line indicates the image derivates from the center. See Figure 1 and
Figure 2.
Figure 1:correct position (expert mode)
Figure 2:incorrect position (expert mode)
In easy mode, move the bladder image to the center. Dark green central line indicates the image is
centered, while the orange central line indicates the image derivates from the center. See Figure 3 and
Figure 4.
Figure 3: correct position (easy mode)
Figure 4: incorrect position (easy mode)
If the central position of the bladder deviates from the central line and to the right of the image, move the
probe to the left. The central position of the bladder moves according to the movements of the probe. If
the bladder image deviates from the central line and to the left side, move the probe to the right.
In both Expert mode and Easy mode, a crosshair of project icon guide is displayed at the right side on the
screen. It helps re-aim and re-scan. (You can enable/disable this function from SETUP
menu>Display>Projection ON/OFF)