CALIBRATION
Please be kindly noted, out of the proprietary technologies in ultrasound imaging and algorithm in
measurement, Caresono PadScan Bladder Scanner HD5 retains clinical supremacy of Zero Calibration
during its lifetime. Regarding this, we hereby officially specify:
Within its lifetime clinical application, under the intact condition of both the PadScan Bladder Scanner
HD5 main unit and the probe:
1.
The operators do not need to do the calibration for clinical measuring precision before their first scan
on the patients.
2.
The operators do not need to do the calibration for clinical measuring precision during their day to
day working.
3.
The operators do not need to do the calibration for clinical measuring precision if changing another
probe.
If the calibration were the fixed procedure from the hospitals or any type of end-user customers, they
could follow the next steps; But please be kindly noted, the phantom model Dansk 616 is the only one
model to comply.
1.
Prepare the Dansk Tissue Equivalent Phantom for
calibration.
2.
Place the probe head into the phantom holder firmly.
3.
Align the probe button with the red arrow on the phantom.
4.
In the main PadScan HD 5 screen, tap SETUP>GENERAL>
CALIBRATION to enter the calibration screen.
5.
Tap
to start auto calibration. It takes about 2
minutes to finish the calibration.
6.
Do not remove the probe during the calibration.
7.
The default calibration value is 32.
8.
A last calibration date and time will be displayed on the