Monthly
CS 9300C User Guide (SM750)_Ed 01
6–3
3D head rest
Nasion support
To clean and disinfect the components:
1. Clean and remove all visible soil by scrubbing with a surfactant, detergent, and water.
2. Rinse with water to remove chemical or detergent residue.
3. Disinfect with an EPA-registered, hospital disinfectant for low-level activity (for
example, quaternary ammonium compounds, some phenolics, and some iodophors).
You must use the disinfectant according to the manufacturer instructions.
Monthly
Wipe the outer covers of the unit with a soft, dry cloth.
Annually
We recommend a general inspection of the unit carried out by an authorized service
technician.
CAUTION
If there is a visible contamination with blood, you must clean
the components and accessories with an EPA-registered
intermediate-level hospital disinfectant that has a
tuberculocidal claim. You must use the disinfectant according
to the manufacturer instructions.