repArAtIon
Before starting a disinfection cycle, make sure that:
A. Lens & Cameras should be capped before exposing to UV-C light.
B. Viewfinders & optics should be concealed by tape or other methods.
C. Do not expose laminated, color filters or photo sensible components to UV-C light.
5.2 t
he
d
osImeter
CArd
CARTONI
recommends the patented UVC 100 Dosimeters from Intellego Technologies.
https://uvcdosimeters.com/uvc-100-dosimeter/
UVC Dosimeters change color when exposed to UV-C energy so you can independently validate your
devices and procedures to see if an optimal dose of UVC has been delivered to a surface.
A recent study from Boston University and Signify has shown that SARS-CoV-2, the virus that causes
COVID-19, can be inactivated by UV-C in 25 seconds with a dose of 22 mJ/cm². The
UV-C LED BOXER
can irradiate an average of 15mJ/cm
2
per minute so you can set the timer to 5 minutes to be in safety
margin. For other types of viruses and bacteria consult the official channels. Having a visual indication of
dose is critical for the success of your UV-C disinfection protocols. UVC Dosimeters are recommended
for use during disinfection cycle to provide visible confirmation and confidence for staff and end users.
https://www.researchsquare.com/article/rs-65742/v1