32
WhEN mEASUrINg ...
Problem
Root Cause
Solution
Abnormal measuring
results:
- Too high; OR
- Too low; OR
- Huge difference
between two recent
measurement.
Incorrect posture
Please step on the platform
barefooted and stand still.
The device is located on the soft
ground such as a carpet OR on
a rugged surface.
Please place the device on a
flat, hard surface.
Cold body that may results
in bad blood circulation.
Warm up your hands and feet to resume
blood circulation and then measure again.
Cold Electrodes.
Place the device in a warm room
for a while and then measure again.
Either your hands or your
feet are too dry.
Wipe your feet with a damp cloth, keeping
them slightly damp when starting measurement.
No display on LCD
when the device
powers on.
Batteries not yet installed.
Install the batteries.
(Please refer to
Insert the Batteries
)
Worn batteries.
Replace all four batteries at the same time.
Please purchase the authorized batteries
for replacement.
CANNOT proceed to
analyze body fat,
total body water,
muscle mass.
Step onto the platform
wearing socks or shoes.
Please keep barefooted during the
measurement, and keep full contact with the
electrodes as well.
The system cannot identify
the possible User ID with most
similar data.
Please assign a User ID following the
instruction in
Set Up Your Profile.
The user fails to select the User
ID from what the system found.
Please assign a User ID following the
instruction in
Set Up Your Profile.
The device powers off
automatically.
Low battery.
Replace all four batteries at the same time.
Please purchase the authorized batteries for
replacement.
ErrOr prOmpT
Error
Description
Solution
Overload.
The device will power off.
Stop using this scale for
measurement.
Low Battery. The device
will power off in four seconds.
Replace all four batteries at the
same time. Please purchase the
authorized batteries for
replacement.
Summary of Contents for BF-1255
Page 1: ...BF 1255 015 09 ...
Page 2: ......
Page 39: ...39 MEMO ...
Page 40: ...40 MEMO ...
Page 41: ......