Reference Manual V1.10
Software Versions 3.xx
X300-621-110
Page 33
It is only possible to circumvent the passcode at the factory.
Care must be taken when setting the Full Digital Setup Passcode to ensure that
the instrument does not become permanently locked.
KEY.LOC
(Front Panel
Key Locking)
This item allows individual keys to be locked and unlocked.
-
The display shows a dash (
–
) to indicate that a key is locked (inactive) or
characters for each key that is active (ie. the characters
P12345
display).
-
The letter
P
represents the
<POWER>
key while the numbers
1234
and
5
represent the remaining operation keys.
-
The operation keys are numbered from the left with the
<ZERO>
key
being number
1
.
Note:
When the
<POWER>
key is locked, the instrument cannot be turned off
from the front keypad.
Default: P12345 - All keys are unlocked (active)
KEY.FN
(Key
Functions)
The function of the keypad
<FUNCTION>
key can be selected here. Refer to
Special Functions page 46 for details of the available key functions.
Options are:
NONE
TEST
COUNT UNITS
HOLD
PEAK.H SHOW.T HI.RES A.TARE CHECK.W
Default: NONE
AUT.OFF
(Auto Power
Off / Battery
Operation)
The instrument can be set up to automatically power down after a period of no
activity. Weight motion, network communications or any press of the keyboard
is enough to keep the instrument powered on. When operating on batteries the
instrument will turn off after 30 minutes of inactivity even if set to Auto.
Options are:
NEVER: Never power off automatically1, 5,10, 30, Auto (time in minutes)
Default: Auto
B.LIGHT
(Backlight
Operation)
Sets the operation of the backlight. When operating the backlight with batteries
the brightness is lowered automatically to conserve power and the backlight will
automatically turn off after 10 seconds of inactivity. To turn on again, press the
<POWER>
key.
Options are:
OFF: Backlight is off.
ON: Backlight is on when weight motion, network communications or any
keypress is detected.
FORCED: Backlight is permanently on.
Default: ON
REM.FN
(Remote
Function)
This item allows the indicator to be triggered from a remote input (see section
3.9.2 Remote Input). The remote input can be set to have:
-
no function (ie. NONE) or
-
it can be set to mimic one of the front five panel key functions (ie. where
KEY1=ZERO, KEY2=TARE, KEY3=GROSS/NET, KEY4=PRINT and
KEY5=FUNCTION).
The remote input can also be set to BLANK where, when the remote input is
pressed and held, the indicator displays dashes (ie.
- - - - - -
) and the front five
key functions are disabled. When the remote input is released the indicator
display and front key functions return to normal.
For the REM.FN to have effect, the SERIAL:TYPE item must be set to
AUTO.