8.
Setup
The instrument digital setup facilities provide the means to configure and calibrate the instrument.
8.1.
Accessing Setup
There are two methods to access the Setup area. For further details of menu items available
in each setup mode, refer to the Setup Menu Quick Reference page 51.
8.1.1. Access using Full Setup
The
Full Setup
method provides access to all functions in Setup, including legal for trade and
calibration sensitive settings. Changes in Full Setup mode may result in the calibration
counter being incremented. Items marked with
indicate that the setting is trade critical.
Changes to passcodes and restoring the factory default settings can only be accessed in Full
Setup mode. These items will however not increment the calibration counter. If an attempt is
made to enter Full Setup using the incorrect passcode, the instrument will respond with the
message
ENTRY DENIED
. Refer to Passcodes page 25 for more information.
Full Setup
1. Instrument on.
2. Press and hold both the
<POWER>
and
<GROSS/NET>
keys together for two seconds.
3.
Full Setup
can also be entered by pressing a push-button accessed from the rear of the
instrument. Refer to OPTIONS:R.ENTRY page 31 for more details.
WARNING
All items in all menus will be enabled in
Full Setup
.
Care should be taken to avoid inadvertently altering the Build or Calibration settings.
8.1.2. Access using Safe Setup
The
Safe Setup
method restricts access to the Trade Critical settings. Changes made in this
mode will not increment the calibration counter. In this manual, items marked with
indicate
that the setting is trade critical. If an attempt is made to enter Safe Setup using the incorrect
passcode, or if an attempt is made to alter a trade critical setting while in Safe Setup, the
instrument will respond with the message
ENTRY DENIED
. Refer to Passcodes page 25 for
more information.
Safe Setup
1. Instrument on.
2. Press and hold both the
<POWER>
and
<TARE>
keys together for two seconds.