4.4 Turning On the Power
1
Insert the AC adapter plug in to the DC IN
connector
on the back of the balance. Place the AC
adapter cable as shown in figure and hold
it with the adapter cable holder.
2
Plug the AC adapter to the power outlet.
After the balance performs a self check,
calibration will be automatically executed.
During this process, the display will change
as follows. “CHE 5”, “CHE
4” ...... “CHE 0”, “CAL 2”, “CAL 1”,
“CAL 0”, “CALEnd”,
“
oFF”
(This span calibration immediately
after power up can be aborted by
pressing the
[POWER]
key, however at
least one span calibration is necessary
before use. The CAUY series does not
perform calibration and the display
showing “CAL...” will not appear.)
A power cable may be necessary to connect the AC adapter to the power
source, depending on the type of the AC adapter
3
Press the
[POWER]
key. After all displays appear, the gram (g) display appears. Also,
the display backlight illuminates (CAUW series only).
4
Pressing the
[POWER]
key again turns on the standby symbol and puts the balance in
standby (warm up) status. (For CAUW-D/CAUW/CAUX series, the current time is
displayed according to the built-in clock settings in 8. For the CAUW series, the backlight
will turn off.)