10.3 Calibration Settings
One of the following four calibration methods listed in the below step 3 is selected as “preset
calibration method”. Preset calibration method will be called with only one key touch from
the mass display for convenience of frequent use. Select your most frequently used method
here. The default setting is “i-CAL” (“E-CAL” for CAUY). The rest of the methods are
alternative calibration methods, which can also be performed at any time by entering the
function selection menu. Procedures for performing each calibration method as the preset or
as an alternative calibration method are described in 10.2.
10.3.1 Selecting Preset Calibration Method
1
From the mass display, press the [CAL] key
repeatedly until the “SEttinG” menu appears.
Press the [O/T] key. “CAL dEF” appears.
2
When “CAL dEF” appears, press the [O/T] key.
The “E CAL” display appear.
3
Press the [CAL] key repeatedly until the
desired setting appears. Select from the following
four types (or two types for CAUY series).
The stability mark appears when the currently
preset method is shown.
“E CAL”: Span Calibration with external weights
(E-CAL)
“E tESt”:
Calibration test with external weights
(E-tESt)
“i CAL”:
Span Calibration with built-in calibration
weight (i-CAL)
(CAUW-D/CAUW/CAUX series only)
“i tESt”: Calibration test with built-in calibration
weight (i-tESt)
(CAUW-D/CAUW/CAUX series only)
4
Press the [O/T] key when the desired calibration
setting appears. “CAL dEF” appears when
the setting is made.
5
Press the [POWER] key to return. The calibration
type set here can be executed by simply
pressing the [CAL] key followed by the [O/T]
key from mass display.
(Displays of four menu items)