12.2 Percentage (%) Conversion
Setting a standard sample to 100% allows percentage conversions.
1
Register percentage unit beforehand. (See
12.1.) Percentage unit is registered by default.
Therefore, registrantion is not required if the
default setting has not been changed.
2
From the mass display, press the
[UNIT]
key
repeatedly to switch to the % display.
3
When a weighing vessel (container) is used,
place the vessel on the pan and press the
[O/T]
key. (The display does not change.)
4
Load the item to serve as a standard (the item
mass serves as 100% and must be at least 100
counts in gram display), and wait for the
stability mark to appear.
5
Press the
[CAL]
key. After “SEt” is displayed,
the mass of the standard item is set as 100%.
6
Remove the standard item and begin
measurement by loading samples.
7
To return to the mass display, press the
[UNIT]
key several times.
x
When the
[UNIT]
key is pressed, the display returns to the mass display, with
units such as g. The actual mass of the current sample can easily be
displayed at any time this way.
x
When changing the item to serve as a 100% standard, the process must be
repeated from step 3.
x
In steps 2 through 4, the display is output in percentage based on the last
setting of percentage conversion.