79
4.4.2 Error Display
Error display
Description
Countermeasure
The load on the pan is unstable at calibration.
Only one of the calibration levers is working.
Avoid wind and vibration.
The drift of zero point is large at calibration.
Empty the pan.
The drift is large at the time of PCAL.
Use correct weight.
The drift is large at span calibration.
Use correct weight.
Weight is wrong.
Use correct weight.
x
Stops when this is displayed.
*
Received command code is not correct.
Check delimiter etc.
Integer of the displayed unit exceeded 7 digits.
Decrease the load.
x
Failure in the balance.
*
PCAL password error.
“___“ is displayed after the error.
Check the password.
Press the
key and
return the balance to mass
display mode.
Attempted to set improper numeric value.
Review the numeric value
and decimal places.
Necessary conditions and numeric values are not
set.
Check graphic display mode.
key was pressed during menu lock.
Release menu lock.
Mode for operation is different.
Check graphic display mode.
Cannot be stored, as the supply voltage is
abnormal.
Check the supply voltage.
*Contact your CAS representative.
Summary of Contents for CBW