43
Note 1. Command Mode Table
Data Request Signal of CI-200
0
1 2 3 4
5
6
7
8
9
10
11
Descriptions on
Request Signal
CI-200
Output Signal
D
dd
K Z CR
LF
Zero Point Key
Received Data Return
D
dd
K T CR
LF
Zero Point Key
Received Data Return
D
dd
K G CR
LF
Gross Weight Key
Received Data Return
D
dd
K N CR
LF
Net Weight Key
Received Data Return
D
dd
H
D
CR
LF
Hold
Key
Received Data Return
D
dd
K B CR
LF
Print Key
Received Data Return
D
dd
K C CR
LF
Total Print Key
Received Data Return
D
dd
K W CR
LF
Weight Data Request Signal Received Data Return
D
dd
I
D
0 0 0
0 0
CR LF
Device
Number
Received Data Return
D
dd
H
Y
0 0 0
0 0
CR
LF
Key Tare Value
Received Data Return
D
dd
H I
0
0
0
0
0 CR
LF
High Limit(LCD Only)
Received Data Return
D
dd
H L
0
0
0
0
0 CR
LF
Low Limit (LCD Only)
Received Data Return
Note 1. (D : 0x44, dd:00~99, K:0x4B , Z:0x5A , CR : 0×0D, LF: 0×0A)
dd = Device Number (2byte), CR = 0×0D, LF: 0×0A
Ex) If a device number is 10, dd becomes 0x31 and 0x30.
Ex) If you want to operate the zero point key when a device number is 11, the indicator operates zeroing if the hex
code of “ 44 31 31 4B 5A 0D 0A” is sent.
Summary of Contents for CI-200 Series
Page 1: ...2 ...
Page 11: ...12 2 Specifications in Appearance 2 1 External Dimension CI 200A CI 201A ...
Page 12: ...13 CI 200S CI 200SC ...
Page 13: ...14 2 2 Front Panel Descriptions CI 200A CI 201A ...
Page 15: ...16 CI 200S CI 200SC ...
Page 78: ...79 MEMO ...
Page 79: ...80 MEMO ...
Page 80: ...81 MEMO ...
Page 81: ...82 ...
Page 82: ...83 ...