25
CAS ELECTRICS CO.,LTD
4.1.1 Span Calibration (Menu Code 8110)
(Calibration MENU -> 1. Calibration -> 1. Span Calibration)
* Requires set of certified weights. (For best result prepare 15kg/6kg(max) weights)
Display shows the amount of weight that you will need.
Caution:
Span calibration
must take place after “
Capacity & Units
” setting.
① Select “Span Calibration”
② Empty tray and press [ENTER]
While calibrating zero display shows “Wait4” ~ “Wait0” and follow next message for Span Calibration.
③ Put on the Weight for Max. Capacities then press [ENTER]