14
2.2 Key Panel and Operation
Functions of the keys
During Weighing
Key
Press Once and Release
Press and Hold for About 3 Seconds
[POWER]
Switches between the operation and standby
modes.
Exits the application function and returns to
the mass display.
[CAL]
Enters span calibration or menu item
selection. (*1)
Displays the last menu item that was set.
(Last menu recall)
[O/T]
Tares the balance. (Displays zero.)(*2) (*5)
Displays the Pretare value.
[UNIT]
Changes the weighing unit or selects
specific gravity measurement. (*3)
Switches between the 1d and 10d
display. (*4)
[PRINT]
Sends the displayed value to a peripheral
device.
Sends the date and time to a peripheral
device.
*
1 This key is used to set values when percent (%), number (PCS), solid specific gravity
(??d), or liquid specific gravity (
▼
d) are displayed.
*
2 When a Pretare value is set, zero is not displayed and [- Pretare value] is displayed.
*
3 Units other than “g” must be set up before they can be used for measurement. Only gram
(g), percent (%), and piece counting (PCS) are set-up before shipment. To set up other
units or specific gravity measurement, refer to section 12., or 14.1, 14.2.
*
4 When the unit is set to 10d, the resolution of the minimum display is decreased by one
decimal place.
*
5 In Pouring mode (See 11.2), the right-most part of
[O/T]
key marked with a circle
functions as the switch for environmental condition setting. Otherwise this part functions
the same way as the other parts of
[O/T]
key.
During Weighing
Key
Press Once and Release
Press and Hold for About 3 Seconds
[POWER]
Returns to the previous menu level
Returns to the mass display.
[CAL]
Moves to the next menu item.
Displays the last menu item that was set.
(Last Menu Recall)
[O/T]
Selects or sets the currently displayed menu item,
or enter into the displayed menu.
No operation.
[UNIT]
Increases the numeric value of the blinking digit by 1.
No operation.
[PRINT]
Moves to the next digit during numeric value entry.
No operation.
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...