13.4 Auto Zero
When the displayed value is within the Zero Range and stability is detected, zeroing occurs
automatically.
The Zero symbol appears in the display when the Auto Zero function is active.
Other keys function as expected with the Auto Zero function activated.
Select menu item
41
to activate it.
x
See 13.5 for details of Zero Range.
x
See 11.3 for details of stability detection.
x
Pretaring Value (13.7) function cannot be used with Auto Zero function.
x
Auto Zero setting can be cleared without menu operation, by pressing
the
[POWER]
key for about three seconds.
x
Not more than one of these seven functions can be set at a time. Setting the second
one with the menu will automatically clear the previously set function; Auto
Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval Timer(14.4),
Add-on mode(14.5), Animal Weighing mode(14.6), Formulation mode(14.7).
x
When minimum display has been switched to 10d (See 9.2), the optimized
Zero Range for 1d display may cause failure in the operation.
Select optimal Zero Range value again when minimum display has been changed
(See 13.5 for details).
Summary of Contents for CU series
Page 1: ...2 ...
Page 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Page 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Page 20: ...21 c Small pan model minimum display 0 001g ...
Page 106: ...107 ...
Page 107: ...108 ...
Page 108: ...109 ...
Page 109: ...110 ...
Page 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Page 124: ...125 MEMO ...
Page 125: ...126 MEMO ...
Page 126: ...127 ...
Page 127: ...128 ...