EB
Service Manual
15
2007/06/13
(3) Press [TARE] key, and then “ G-1“ message and “9.7994” will be shown. The first digit,”9”
will blink.
(4) Input a gravitational acceleration value by using [ZERO] key.
(5) Press [TARE] key, and then “G-2“ message blinks.“9.7994” will be shown. The first
digit,”9” will blink.
(6) Input a gravitational acceleration value by using [ZERO] key.
(7) Press [TARE] key to save the gravitational acceleration value, and “C-9 ” message will be
shown.
2.1.4. Calibration factor Setting (C-10)
(1) Pressing and holding “Calibration Switch” press [POWER] key.
(2) After “CAL” message blinks three times and shows the version of scale, it displays “CAL-1”
message.
(3) Press [ZERO] to display “C-10”.
(4) Press [TARE] key, and then “UNIT “ message and “0” will be shown. The first digit,”0”
will blink. It means calibration unit is “kg” (0 : kg, 1 : lb)
(5) Input a calibration unit by using [ZERO] key.
(6) Press [TARE] key, and then “CAPA“ message blinks.“0015” will be shown. The first digit,”0”
will blink. It means a full-capability is “15 (calibration unit, kg or lb)”
(7) Input a capability by using [ZERO] key.
(8) Press [TARE] key, and then “MCAPA“ message blinks.“0005” will be shown. The first
digit,”0” will blink. It means a mid-capability is “05 (calibration unit, kg or lb)”
(9) Input a capability by using [ZERO] key.
(10)Press [TARE] key, and then “W-dP “ message blinks.“3” will be shown. The first digit,”3”
will blink. It means a weight decimal point is “3 (will display 0.000)”
(11)Input a weight decimal point by using [ZERO] key.
(12) Press [TARE] key, and then “1d “ message blinks.“0.005” will be shown. The third digit,”0”
will blink. It means a division is “0.005 (calibration unit, kg or lb)”
(13) Input a division by using [ZERO] key.
(14) Press [TARE] key, and then “dual “ message blinks.“1” will be shown. The third digit,”1”
will blink. It means a dual interval is disable. (0 : disable, 1 : enable)”
(15) Input a dual interval enable by using [ZERO] key.
(16) Press [TARE] key to save the calibration factor, and “C-10 ” message will be shown.
2.1.5. Displaying Real A/D Value (C-5)
Display Raw AD
Summary of Contents for EB Series
Page 1: ...EB Service Manual 1 2007 06 13 EB SERVICE MANUAL ...
Page 6: ...EB Service Manual 6 2007 06 13 1 4 Dimension ...
Page 9: ...EB Service Manual 9 2007 06 13 1 6 Sealing Method PLATE ...
Page 10: ...EB Service Manual 10 2007 06 13 STICKER ...
Page 22: ...EB Service Manual 22 2007 06 13 4 Exploded View ...
Page 23: ...EB Service Manual 23 2007 06 13 5 Load Cell drawing ...
Page 24: ...EB Service Manual 24 2007 06 13 6 Part Location 6 1 Main PCB Top ...
Page 25: ...EB Service Manual 25 2007 06 13 6 2 Main PCB Bottom ...
Page 26: ...EB Service Manual 26 2007 06 13 6 3 Rear Display PCB Top 6 4 Rear Display PCB Bottom ...
Page 27: ...EB Service Manual 27 2007 06 13 6 5 Terminal PCB Top 6 6 Terminal PCB Bottom ...
Page 28: ...EB Service Manual 28 2007 06 13 6 7 Cal PCB Top ...