6 Calibration Mode
20
< CAL: MAIN MENU >
1. CALIBRATION
5. NETWORK OPTIONS
2. SYSTEM OPTIONS
6. SELF TEST
3. DIGITAL FILTERING 7. NORMAL FUNCTION
4. PRINTER SETTINGS 8. PRICE OPTIONS
6
Calibration Mode
6.1
Entering Calibration Mode
1.
Make sure that the Main Power is OFF (switch on left side of scale.)
2.
Press and hold down the ON/OFF key on the upper right corner of the keyboard.
3.
While you are holding down the ON/OFF key, turn the Main power ON.
4.
Once you here a series of “chirps” release the ON/OFF key.
5.
After a few seconds the scale will be in CAL mode.
The CAL Main Menu
screen will appear as follows:
Pressing ESC from this
menu will take you out of CAL
mode and affect all of the
changes that you may have
made.
NOTE: Only CAS trained personnel should attempt to make changes in CAL mode. If you are not
trained to work on this equipment, please contact the CAS (USA) Service Department for assistance.
Non-qualified personnel attempting service the CAS LP-2, risk void the scale’s warrantee.
6.1A Entering Calibration Mode (alternate method)
Some units have a working calibration switch located on the top of the upper case, underneath
the platter. Although USA models may have CAL switches in this location, these switches do not have
any effect. If your scale has a working CAL switch, then you must seal the scale physically.
ALL OF THE SCREENS TO FOLLOW SHOW USA DEFAULT SETTINGS.