50
10. ERROR MESSAGE
Err 01
■
Cause
The Indicator can’t be initialized due to platform shaking
■
Management
Check the A/D value (at ‘TEST 3’) .I
If the load cell is damaged, replace the load cell
Put the platform on a flat place without shaking and power on.
Err 02
■
Cause
Initial A/D value of zero is too high or low, when the power is ON
Or in the case of NC-I is lying on the flat
■
Management
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Check that the load cell is well connected with main unit
Or calibration is done wrongly
Err 09 (Zero Key)
■
Cause
The current weight exceeds zero range (2 % of Maximum Capacity)
■
Management
Unload the scale
Err 10 (at Zero calibration)
■
Cause
Initial A/D value of zero is too high or low.
■
Management
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Check that the load cell is well connected with main unit
Replace the loadcell.
Summary of Contents for NC-1
Page 1: ...2 ...
Page 20: ...21 11 Select hex file 12 Download is done 13 If download is done you must turn off SW5 FIG 2 ...
Page 24: ...4 Cal schematic SCHEMATIC DIAGRAM 25 ...
Page 25: ...26 8 PCB BOARD DIAGRAM 1 Main PCB board LED ...
Page 26: ...27 2 Display PCB board ...
Page 27: ...28 3 Charger PCB board ...
Page 28: ...29 9 DEVICE OF SPECIFICATION 1 MPU ADUC845 ...
Page 29: ...30 ...
Page 30: ...31 ...
Page 31: ...32 ...
Page 32: ...33 2 OPAMP OP2177 OP2277 ...
Page 33: ...34 ...
Page 34: ...35 ...
Page 35: ...36 3 REGULATOR LM2575 ...
Page 36: ...37 ...
Page 37: ...38 ...
Page 38: ...39 ...
Page 39: ...40 ...
Page 40: ...41 ...
Page 41: ...42 ...
Page 42: ...43 4 RESET TC1274 ...
Page 43: ...44 ...
Page 44: ...45 ...
Page 45: ...46 ...
Page 46: ...47 4 RESET TC1274 ...
Page 47: ...48 ...
Page 48: ...49 ...
Page 82: ...83 12 EXPLODE VIEW ...
Page 83: ...84 ...