12.Error Message & Trouble Shooting
Error
Reason
Solution
Err 01 Initial value of A/D is fail
Check the L/C connector & test
A/D conversion in Test Mode 2
Err 02
Load cell connection failure,
Error in A/D conversion part
Check the L/C connector & test
A/D conversion in Test Mode 2
Err 05 Input value is over the range
Input the value from 32 to 255
Err 07 Product ID is larger than 19
The range of Product ID is 0 to 19
Err 08
Hi limit weight is larger than
Maximum weight
Set the Hi limit weight to under
the maximum weight
Err 09
Low limit weight is larger
than hi limit weight
Set the Low limit weight to under
the Hi limit weight
Err 11
The unit of weight is different
Can’t process cause the unit of
weight calibration is not same
Err 13
Zero value exceeds the initial
zero range
Remove a weight from the platter
and turn on power
Err 21 The resolution exceeds 1/20,000
Lower the resolution
Err 22
he weight for span calibration
is set to under 10% of maximum
capacity
The weight of Span calibration
Should be set to over 10% of max.
Capacity (Check it in CAL 1)
Err 23
The weight for span calibration
is set to over 100% of maximum
capacity
The weight of Span calibration
Should be set to under 100% of
max. capacity (Check it in CAL 1)
Err 24 The value of Span is high or low
Adjust resolution
Err 31
The password before and the cur-
rent password are not same
Input the new password again
Err 32 The number of sample is too low
Put more samples or
Decrease the number of samples
Err 50 Internal memory data is erased
The product is in trouble
Contact the A/S center