INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
These units are built for use indoors at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
at
no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating outside these conditions may
adversely affect the unit temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can adversely
impact its temperature performance. These include:
•
Proximity to other ovens, autoclaves, or any other device producing significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
R
EQUIRED
C
LEARANCES
These clearances are required to provide air flows for ventilation and cooling.
6 inches (152 mm)
of clearance is required on the sides.
24 inches (610 mm)
of headspace clearance is required between the exhaust vent and any
overhead cover or partition.
o
12 inches (305 mm)
of vertical headspace clearance will suffice if the oven exhaust is
vented from the workspace through a duct or other channeling.
Do not place objects on top of the oven.
Allow at least
6 inches (152 mm)
from the access port and fan vent on the back of the oven to the
nearest wall or partition. Keep the fan unobstructed at all times.
6” (152 mm) Fan
6” (152 mm)
6” (152 mm) 12” (305 mm)
24” (610 mm)
24” (610 mm)
Door Swing 32”
(813mm)
Access Port
12” (305 mm) if
ducting is attached
Summary of Contents for CDO-5-2
Page 1: ...Installation Operation Manual CDO 5 2 Drying Decarboxylation Oven 220 240 Voltage...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 10: ...10 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 44: ......