USER MAINTENANCE
Continued next page
Calibration continued
3
Place the oven in temperature calibration mode.
a.
Press and hold
both
the
UP
and
DOWN
arrow
buttons simultaneously.
•
The Temperature Display will show the letters “C O”,
then begin flashing the
current temperature value
.
Note:
If an arrow key is not pressed for five seconds, the
Temperature Display will cease flashing, and store the last
displayed value as the new current chamber temperature
value.
Set Temperature
Set Temperature
Current Temp. Value
4
Adjust the current temperature value to match the reference
device.
a.
Use the
UP
and
DOWN
arrow buttons.
Set Temperature
Corrected
Reference Device
5
After entering the correction adjustment, wait 5 seconds.
•
The temperature display will cease flashing and
store the correction as an offset.
•
The oven will now begin heating or allow itself to
cool in order to reach your setpoint with the
corrected display value.
Set Temperature
Heating with Corrected
Value
6
Wait for 30 minutes for the oven to stabilize
after the
oven has achieved the setpoint
with the corrected
display adjustment.
•
Failure to wait until the unit is fully stabilized
will result in an inaccurate oven display
reading.
Set Temperature
Summary of Contents for CDO-5-2
Page 1: ...Installation Operation Manual CDO 5 2 Drying Decarboxylation Oven 220 240 Voltage...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 10: ...10 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 44: ......