INSTALLATION
Cabinet Vacuum and Vent Ports
Port and Power Specifications
•
Cabinet Vacuum Port – KF-25
o
Opened and closed using the Vac control on the oven front.
•
Cabinet Power Outlet – NEMA 6-20R
o
Intended to power a vacuum pump.
o
Turned on and off using the Vacuum power switch on the front control panel.
•
Cabinet Intake Vent Port (backfill inlet) – 1/2 Inch (12.7 mm) OD
o
A clean or inert gas supply source may be connected to this port.
o
The maximum allowed gas pressure is 15 psi.
o
Opened and closed using the Vent handle on the oven front.
•
Vacuum Port, Back of Oven – KF-25 Fitting
o
Intended for introducing temperature sensor probes into the oven chamber. Probes
must be inserted through a vacuum-rated feedthrough and secured in the oven
prior to placing the chamber under vacuum.
o
A vacuum pump can be connected to the KF-25 port for increased efficiency in
vacuuming down the chamber. When using this port,
the Vac valve control on the
front panel must be set to closed.
Intake Vent Port
Vacuum Port
Power Outlet
Cabinet Front Panel Removed
Mounting Plate
: The rectangular plate
located on center of the cabinet floor
may be removed, drilled with studs, and
then reinstalled. Used for mounting a
vacuum pump on.
Summary of Contents for CVAC-9.3
Page 1: ...Installation Operation Manual CVAC 9 3 Vacuum Oven ...
Page 5: ...5 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS ...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION ...
Page 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Page 34: ...34 P a g e cascadesciences com Tel 503 847 9047 CONTROL OVERVIEW ...
Page 46: ...46 P a g e cascadesciences com Tel 503 847 9047 OPERATION ...
Page 54: ......