Test the high limit system once per year for functionality.
The high temperature limit is set by the end-user at least 10°C above the highest temperature the
oven will run at during your recipe profile or constant-temperature application.
End of Procedure
1.
Advance to the Limit High Set Point, starting on the Homepage
Push Advance
Multiple Times
•
Push the Advance button repeatedly until “Lh.S1”
(Limit High Set Point) shows in the green mid-
level display line.
2.
Adjust the high limit to at least 10°C above the highest temperature of your application
Adjust
•
The oven will automatically save and apply the
new High Limit setting after you have stopped
adjusting.
Note
: If you are just checking the current high
temperature limit setting, push the Reset button to
exit the High Set Point menu and return to the
homepage without saving any changes.
3.
Return to the Homepage
Push Reset
•
Returned to Homepage
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......