USER MAINTENANCE
Continued on next page
Calibration continued
3
Unlock the controller.
a.
See the Unlocking procedure on page 55.
Note:
The temperature controller must be unlocked in order to
access the Operations menu and enter a calibration offset.
4
Navigate to the Operations menu
after
unlocking the
controller.
a.
Press and hold both the
Up
and
Down
Arrow buttons
simultaneously for approximately 5 seconds.
b.
Release the buttons when “A1” appears on the top
display line and “oPEr” appears in the mid display line.
Operations Menu
5
Advance through the Operations menu options to the
Temperature Calibration offset parameter.
Push the green
Advance button
repeatedly until “i.CA”
appears in the green mid display line and a number value in
the red top line.
6
Adjust the number value in the top display line to match the
offset value from step 2, using the arrow buttons.
7
Save the calibration offset and return to the homepage.
a.
Push the Reset button 3 times so the display shows the
homepage.
•
The oven will begin to heat or passively cool to reach the
current set point using the offset display value for the current
temperature.
x2
x3
Summary of Contents for CVO-2-HT
Page 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Page 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Page 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 72: ......