INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, verify that each of the following requirements is satisfied:
Power Source
The wall power outlet must meet the power requirements listed on the unit data plate.
Model
AC Voltage
Amperage
Frequency
CVO-2
110 – 120
6.0
50/60 Hz
•
The unit may be damaged if the supplied voltage varies by more than 10% from
the data plate rating.
•
Wall power sources must be
single phase (1)
and
protective earth grounded
.
•
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure. The
circuit must meet or exceed the amperage requirements listed on the unit data plate.
•
The recommended wall circuit breakers for these units are:
o
CVO-2:
15 amps
•
The wall power source must conform to all national and local electrical codes.
Power Cords
The unit must be positioned so that all operators can quickly unplug the power cord in the event of
an emergency.
•
The
CVO-2
is provided with a
125V, 15A, 9ft 5in (2.86m), NEMA 5-15P
power cord.
•
Always use the same type of power cord shipped with the oven.
Continued next page
CVO-2
Standard NEMA
5-15R Outlet
Summary of Contents for CVO-2
Page 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 52: ......