INSTALLATION
C
ONNECT TO A
G
AS
B
ACKFILL
S
UPPLY
Optional:
A clean or inert gas supply — such as nitrogen (N
2
) or argon — may be connected to the
vent intake port located on the back of the oven. This supply will backfill the evacuated oven
chamber when the vent intake valve is opened on the front control panel.
The maximum allowed gas backfill pressure is 15 psi.
Exceeding 15 psi of gas backfill
pressure may damage the oven.
CVO-2
Gas Supply
Intake Vent – To the oven
chamber ¼ in (6.35mm) OD