CONTROL OVERVIEW
Light Switch
Controls the oven chamber light.
Power Switch
Controls all power to the oven and its systems.
Set Temperature Display and Controls
Shows the current chamber temperature. The
Up
and
Down
arrow buttons are used to access the
Temperature Set Point (SP) or Calibration Offset (C O) display modes and input the temperature set
point or calibration adjustment value.
Hold Button
Temporarily disables heating in the oven chamber. Pushing the button once turns off power to the
heating elements. Pushing the button a second time restores heating.
Over Temperature Activated
This red light illuminates if the Over Temperature Limit heating cutoff system routes power away
from the heating elements. The heating cutoff is triggered when the shelving temperature exceeds
250°F. For more details, please see the
description on page 38.