C
OOLDOWNS
The oven chamber is well insulated and requires a significant length of time to cool down while
remaining sealed.
•
6 hours may be required to cool down from 150°C to 100°C.
•
16 hours may be needed for the oven to return to room temperature cooling from 150°C.
•
Backfilling the oven does not significantly increase the rate of cooling.
Introducing free atmosphere into the oven chamber at temperatures above 100°C risks oxidizing
chamber surfaces.