OPERATION
P
UMP
D
OWN AND
B
ACKFILL
–
S
IDE
V
ALVE
Reminder:
Verify which valve control your process uses for backfilling the oven (Backfill or Vent).
End of Procedure
Pump Down the Oven Chamber
1.
Verify the Vacuum, Vent, and Backfill Valve controls are in the closed position
•
This protects your vacuum pump from exposure to
streaming atmosphere.
2.
Turn on your vacuum pump
•
If using a building supply system, verify it is providing
vacuum to the oven vacuum port.
3.
Open the oven side Vacuum Valve
Turn the control all the way counterclockwise.
•
The Vacuum level on the main controller
should show the chamber pressure
decreasing.
•
The achievable vacuum level is dependent on altitude
above sea level as well as the vacuum supply efficiency
and the volume of outgassed byproducts.
Holding at Vacuum
Continue evacuating the chamber throughout the baking
application to vent outgassed byproducts.
Backfilling the Oven Chamber
4.
Close the side Vacuum Valve
Turn the side Vacuum Valve control back to the closed position
(clockwise) to protect the vacuum pump from extended exposure
to streaming atmosphere.
•
The pump may remain on.
5.
Slowly open the Vent or Backfill valve per your SOP
•
The chamber will return to room pressure.
VACUUM
VENT or Backfill
VACUUM
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......