21
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INTRODUCTION
H
AZARDOUS
Z
ONES
M
AP
Zone 1:
Explosive concentrations of gases or vapors are frequently present during normal operations
inside
the Zone 1 spaces of the oven chamber, KF-25 fitting, and vacuum system.
Zone 2:
Any explosive concentrations of gases or vapors in the Zone 2 space surrounding the
oven, its KF-25 fitting, and the vacuum system are unintentional, infrequent, and short-lived –
typically the result of an accident, a hazardous materials leak, or a ventilation system failure.
Explosive concentrations are never present as a part of normal operations in the Zone 2 space.
Zone 2 Hazardous Zone Assessment:
The full extent of the Zone 2 hazard area around the
oven and vacuum system must be based on an assessment of
your site
. Assessment factors
include the site airflows, air exchange rate, and the characteristics of your HVAC or other climate
control system.
Zone 1
Zone 2
Map Key
Zone 2 extends approximately 3 feet (1 meter)
from the oven door. The Zone may extend farther
depending on the ventilation characteristics of your
workspace.
The KF-25 Fitting is a Zone 1 space.
Vacuum line connecting to the vacuum pump.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......