50 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
P
LUG
-I
N
P
OWER
S
OURCE
R
EQUIREMENTS
For ovens with plug-in power cords.
When selecting a location for a plug-in oven, verify
each of the following requirements is satisfied:
Power Source:
The power source must meet the power requirements listed on the unit data
plate.
Model
Voltage
Amps
Frequency
CVO-5-EX-SS
115
12
50/60 Hz
CVO-5-EX-2-SS
230
6
50/60 Hz
•
Single Phase and Grounded:
The wall power source must be single-phase (1) and
protective earth grounded.
o
If the unit is not grounded properly, parts such as knobs and controls can conduct
electricity and cause serious injury.
•
Electrical Codes
: The wall power source must conform to all national and local electrical
codes.
•
Safe Voltage Range:
The unit may be damaged if the supplied voltage varies by more
than 10% from the data plate rating.
o
The unit is safety rated to withstand transient overvoltage levels up to Overvoltage
Category II.
•
Separate Circuit
: Use a separate circuit to prevent loss of the unit due to overloading or
circuit failure. The circuit must meet or exceed the amperage requirements listed on the
unit data plate.
Model
Part ID
Power Outlet Type
CVO-5-EX-SS
CSVXBS522-N
Standard
NEMA 5-15R
CVO-5-EX-2-SS
CSVXBS522-NA
NEMA 6-15R
CVO-5-EX-2-SS
CSVXBS522-NE
CEE7 sockets
compatible with
CEE7/7 plugs
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......