INSTALLATION
Power Cord Oven Requirements Continued
Over Current Protection
: A switch or circuit-breaker must be used in the building installation to
protect against overcurrent conditions.
•
Outside Zone 2
: The switch or circuit breaker must be safely located outside the
hazardous location Zone.
•
Circuit Breakers
: The recommended circuit breakers for these units are:
o
CVO-5-EX-SS:
20 amps
o
CVO-5-EX-2-SS:
15 amps
Power Cord
The cord must be plugged in outside of the hazardous Zone to preclude any possibility of sparking
igniting and explosive concentration of gases or vapors.
Do not plug in the cord at this time.
•
Each CVO-5-EX-SS oven comes with an integral
15-foot (4.57-meter)
power cord.
•
The suffix on the end of the oven Manufacturer Part ID indicates which cord type is
attached to the oven.
o
-N
, NEMA 5-15P 115 Volt Cord
o
-NA
, NEMA 6-16P 230 Volt Cord
o
-NE
, Euro CEE 7/7 230 Volt Cord
The power cords are rated to the following specifications:
•
US National Electrical Code permitted for exposed runs outside of sealed conduits in
hazardous locations.
•
Temperature rated to 125°C protection from thermal overloads.
•
Complies with UL 2225 crush and impact requirements for metal-clad hazardous location
cables.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......