55
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
Hardwire Power Requirements Continued
Hardwired Power Cord Disconnect:
The oven must be positioned so that all operators have
access to the power cord disconnect in case of emergencies.
•
The disconnect must be near the equipment and within easy reach of the operator.
•
The disconnect must be marked as the disconnecting device for the equipment.
Over Current Protection
: A switch or circuit-breaker must be used in the building installation to
protect against overcurrent conditions.
•
Outside Zone 2
: The switch or circuit breaker must be safely located outside the
hazardous location Zone.
•
Circuit Breakers
: The recommended circuit breakers for these units are:
o
CVO-5-EX-SS-H:
20 amps
o
CVO-5-EX-2-SS-H:
15 amps
•
Required Safety Compliances
o
Circuit breakers for hardwire (permanent connect) ovens must comply with IEC-
60947-2.
o
Switches must comply with IEC 60947-3.
Power Feed
The power feed is
15 feet (4.57 m)
long and rated to the following specifications:
•
US National Electrical Code permitted for exposed runs outside of sealed conduits in
hazardous locations.
•
Temperature rated to 125°C protection from thermal overloads.
•
Complies with UL 2225 crush and impact requirements for metal-clad hazardous location
cables.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......