64 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
Warning
: The technician must verify there are no hazardous substances or potentially explosive
atmospheres present prior to and while hardwiring an oven to its power supply and when initially testing
the connection.
Avertissemen
t: Le technicien doit vérifier qu'il n'y a pas de substances dangereuses ou d'atmosphères
potentiellement explosives présentes avant et pendant le câblage d'un four à son alimentation électrique.
H
ARDWIRE TO
P
OWER
S
UPPLY
Hardwire Ovens Only
Power Feed Types
Each hardwire oven comes provided with an integral
15 ft (4.57 m)
power feed consisting of:
115 Volt Ovens: CVO-5-EX-SS-H
•
One 14-gauge hot wire: black
•
One 14-gauge earth ground: green-yellow
230 Volt Ovens: CVO-5-EX-2-SS-H
•
One 14-gauge hot wire: black
•
One 14-gauge hot wire: white
•
One 14-gauge earth-ground: green-yellow
Grounding Required
The oven must be earth grounded using the protective conductor terminal
(green with
yellow stripe wire). Do not remove the protective conductor (earth connection). Removing the
protective conductor will negate the oven’s protections against potentially dangerous electric
shocks and create a potential fire hazard.
Connect the Oven Now
A qualified and competent electrical technician should now wire the oven power feed to a power
supply meeting the requirements on page 50. If the connection is being made within the hazardous
area around the oven, the connection must meet all applicable requirements for your jurisdiction to
ensure the connection is isolated from potentially explosive atmospheres. After making the
connection, do not turn on the power supply circuit at this time other than to test the connection.
If you have a plug-in power cord oven, do
not
plug it in now.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......